• Advanced power quality health—At-a-glance power quality health data in real time so you can make better maintenance decisions
  • Energy loss calculator—Discover the costs of energy loss specifically caused by poor power quality
  • Highest safety rating in industry—CAT III 1000 V / CAT IV 600 V

Previous | Next

Power Quality and Energy Analyzers

National Engineering Research and Development Centre


Product Category/ Test Name (Matrix):

Electrical Instuments

Sub Category:



Keywords:





Service Charge (Rs) -









Manufacturer Fluke Funding Source NERDC
Model Fluke 435 Whether the equipment service can be provided to external researchers/institutions Yes
Manufactured Year 0000 Availability of Technically Competent Staff
HS Code Instrument Condition Functioning well
Vendor Name Commencement of Operations Date
Vendor Contact InstrumentPrice 0.00
Vendor Url Stage of the Record Completed
Catalog Aceess
Catalog Upload View File
Catalog Link
Institute National Engineering Research and Development Centre
Faculty
Department Electrical & Electronics Department
Laboratory Electrical and Electronic Lab
Contact Person Name HoD Electrical Department
Email electrical@nerdc.lk
Phone Number 0112236284
Mobile Number
Product Category Electrical Instuments
Sub Category
Testing Paremeters Current, Voltage, Frequency, Power
Description
Relevant Instruments
Technical Specification
Volt
  Model Measurement Range Resolution Accuracy
Vrms (AC + DC) 434-II 1 V to 1000 V phase to neutral 0.1 V ±0.1% of nominal voltage1
435-II 1 V to 1000 V phase to neutral 0.01 V ±0.1% of nominal voltage1
Vpk 1 Vpk to 1400 Vpk 1 V 5% of nominal voltage
Voltage Crest Factor (CF) 1.0 > 2.8 0.01 ±5%
Vrms½ 434-II 1 V to 1000 V phase to neutral 0.1 V ± 1 % of nominal voltage
434-II and 435-II   0.1 V ±0.2% of nominal voltage
Vfund 434-II 1 V to 1000 V phase to neutral 0.1 V ± 0.5 % of nominal voltage
435-II   0.1 V ± 0.1 % of nominal voltage
Amps (accuracy excluding clamp accuracy)
Amps (AC + DC) i430-Flex 1x  5 A to 6000 A 1 A ±0.5% ±5 counts
i430-Flex 10x 0.5 A to 600 A 0.1 A ±0.5% ±5 counts
1mV/A 1x 5 A to 2000 A 1A ±0.5% ±5 counts
1mV/A 10x 0.5 A A to 200 A (AC only) 0.1 A ±0.5% ±5 counts
Apk i430-Flex 8400 Apk 1 Arms ±5%
1mV/A 5500 Apk 1 Arms ±5%
A Crest Factor (CF) 1 to 10 0.01 ±5%
Amps½ i430-Flex 1x 5 A to 6000 A 1 A ±1% ±10 counts
i430-Flex 10x 0.5 A to 600 A 0.1 A ±1% ±10 counts
1mV/A 1x 5 A to 2000 A 1 A ±1% ±10 counts
1mV/A 10x 0.5 A A to 200 A (AC only) 0.1 A ±1% ±10 counts
Afund i430-Flex 1x 5 A to 6000 A 1 A ±0.5% ±5 counts
i430-Flex 10x 0.5 A to 600 A 0.1 A ±0.5% ±5 counts
1mV/A 1x 5 A to 2000 A 1 A ±0.5% ±5 counts
1mV/A 10x 0.5 A A to 200 A (AC only) 0.1 A ±0.5% ±5 counts
Hz
Hz Fluke 434 @ 50 Hz Nominal 42.50 Hz to 57.50 Hz 0.01 Hz ±0.01 Hz
Fluke 434 @ 60 Hz Nominal 51.00 Hz to 69.00 Hz 0.01 Hz ±0.01 Hz
Fluke 435 @ 50 Hz Nominal 42.500 Hz to 57.500 Hz 0.001 Hz ±0.01 Hz
Fluke 435 @ 60 Hz Nominal 51.00 Hz to 69.00 Hz 0.001 Hz ±0.01 Hz
Power
Watts (VA, var) i430-Flex max 6000 MW 0.1 W to 1 MW ±1% ±10 counts
1 mV/A max 2000 MW 0.1 W to 1 MW ±1% ±10 counts
Power Factor (Cos j/DPF) 0 to 1 0.001 ±0.1% @ nominal load conditions
Energy
kWh (kVAh, kvarh) i430-Flex 10x Depends on clamp scaling and V nominal ±1% ±10 counts
Energy Loss i430-Flex 10x Depends on clamp scaling and V nominal ±1% ±10 counts Excluding line resistance accuracy
Harmonics
Harmonic Order (n) DC, 1 to 50 Grouping: Harmonic groups according to IEC 61000-4-7
Inter-Harmonic Order (n) OFF, 1 to 50 Grouping: Harmonic and Interharmonic subgroups according to IEC 61000-4-7
Volts % f 0.0% to 100% 0.1% ±0.1% ±n x 0.1%
r 0.0% to 100% 0.1% ±0.1% ±n x 0.4%
Absolute 0.0 to 1000 V 0.1 V ±5%1
THD 0.0% to 100% 0.1% ±2.5%
Amps % f 0.0% to 100% 0.1% ±0.1% ±n x 0.1%
r 0.0% to 100% 0.1% ±0.1% ±n x 0.4%
Absolute 0.0 to 600 A 0.1 A ±5% ±5 counts
THD 0.0% to 100% 0.1% ±2.5%
Watts % f or r 0.0% to 100% 0.1% ±n x 2%
Absolute Depends on clamp scaling and V nominal ±5% ±n x 2% ±10 counts
THD 0.0% to 100% 0.1% ±5%
Phase Angle -360° to +0° ±n x 1°
Flicker
Plt, Pst, Pst (1 min) Pinst 0.00 to 20.00 0.01 ±5%
Unbalance
Volts % 0.0% to 20.0% 0.1% ±0.1%
Amps % 0.0% to 20.0% 0.1% ±1%
Mains Signaling
Threshold Levels Threshold, limits and signaling duration is programable for two signaling frequencies
Signaling Frequency 60 Hz to 3000 Hz 0.1 Hz  
Relative V % 0% to 100% 0.10% ±0.4%
Absolute V3s (3 second avg.) 0.0 V to 1000 V 0.1 V ±5% of nominal voltage
General Specifications
Case Design Rugged, shock proof with integrated protective holster Drip and dust proof IP51 according to IEC60529 when used in tilt stand position Shock and vibration Shock 30 g, vibration: 3 g sinusoid, random 0.03 g 2 /Hz according to MIL-PRF-28800F Class 2
Display Brightness: 200 cd/m 2 typ. using power adapter, 90 cd/m 2 typical using battery power Size: 127 x 88 mm (153 mm/6.0 in diagonal) LCD Resolution: 320 x 240 pixels Contrast and brightness: user-adjustable, temperature compensated
Memory 8GB SD card (SDHC compliant, FAT32 formatted), up to 32GB optionally. Screen save and multiple data memories for storing data including recordings (dependent on memory size).
Real-Time Clock Time and date stamp for Trend mode, Transient display, System Monitor and event capture
Environmental
Operating Temperature 0°C ~ +40°C; +40°C ~ +50°C excl. battery
Storage Temperature -20°C ~ +60°C
Humidity +10°C ~ +30°C: 95% RH non-condensing
+30°C ~ +40°C: 75% RH non-condensing
+40°C ~ +50°C: 45% RH non-condensing
Maximum Operating Altitude Up to 2,000 m (6666 ft) for CAT IV 600 V, CAT III 1000 V
Up to 3,000 m (10,000 ft) for CAT III 600 V, CAT II 1000 V
Maximum storage altitude 12 km (40,000 ft)
Electro-Magnetic-Compatibility (EMC) EN 61326 (2005-12) for emission and immunity
Interfaces Mini-USB-B, Isolated USB port for PC connectivity SD card slot accessible behind instrument battery
Warranty Three years (parts and labor) on main instrument, one year on accessories
View full family specifications »
1. ±5% if ≥ 1% of nominal voltage ±0.05% of nominal voltage if < 1% of nominal voltage
2. 50Hz/60Hz nominal frequency according to IEC 61000-4-30
3. 400Hz measurements are not supported for Flicker, Mains Signaling and Monitor Mode
4. For nominal voltage 50 V to 500 V
Number of samples per cycle 0
Number of samples per day 0
Total usage hour per day 0


Under Development Stage

Related Instruments