Light Meter displays measurements from LI-COR radiation sensors. The LI-250A provides high-resolution reading and conversion of microamp signals from the attached light sensor into suitable light measurement units. A quantum sensor measures photosynthetically active radiation (µmol s -1 m-2 of PAR), a pyranometer measures solar radiation (W m-2 ), and a photometric sensor measures illuminance (lux).
| Manufacturer | LI-COR |
| Model | LI-250A |
| Manufactured Year | 0000 |
| HS Code | — |
| Instrument Condition | Functioning well |
| Instrument Price (Rs) | 545900.00 |
| Funding Source | HETC/QIG |
| External Research Support | Yes |
| Availability of Technical Staff | Yes |
| Commencement of Operations | 2014-07-21 |
| Vendor Name | Islandwide Scientific (pvt) Ltd |
| Vendor Contact | +94112681150 |
| Vendor URL | — |
| Catalog Access | — |
| Catalog Upload | — |
| Catalog Link | https://www.licor.com/env/products/light/light_meter |
| Stage of the Record | Incomplete |
Please sign in to view the contact details of this instrument and its owner.
Login to View Contact DetailsAccount access is restricted to registered users. Register now for free.
| Testing Service | Light meter |
| Sub Category | — |
| Testing Parameters | Photosynthetically Active Radiation (µmol s -1 m-2 of PAR), Solar Radiation (W m-2 ), Illuminance (lux) |
| Description | Light measurements can be obtain using radiation sensors. The quantum sensor measures Photosynthetically Active Radiation (µmol s -1 m-2 of PAR), The Pyranometer measures solar radiation (W m-2), and the photometric sensor measures illuminance (lux) |
| Relevant Instruments | Light meter, Radiation sensors (quantum sensor, Pyranometer, photometric sensor) |
| Technical Specification | <ul> <li><strong>Accuracy:</strong> <ul> <li><strong>25 °C:</strong> Typically ± 0.4% of reading ± 3 counts on the least significant digit displayed (all ranges).</li> <li><strong>0 - 55 °C:</strong> Typically ± 0.6% of reading ± 3 counts on the least significant digit displayed (all ranges).</li> </ul> </li> <li><strong>Range Selection:</strong> Autoranging (3 ranges).</li> <li><strong>Linearity:</strong> ± 0.05%.</li> <li><strong>Sensors:</strong> Designed for type "SA" sensors; type "SB" and "SL" sensors with BNC connectors are also compatible when calconnector or millivolt adapter are removed.</li> <li><strong>Sensor Calibration:</strong> Each sensor is supplied with a calibration multiplier. Calibration multipliers for two sensors can be stored in memory. Calibration multipliers are entered from the keypad.</li> <li><strong>Signal Averaging:</strong> Sensor output can be collected and displayed as a 15-second average (approximately 60 readings). Averages are retained on the display in HOLD mode.</li> <li><strong>Display:</strong> LCD display. Updated every 0.5 seconds in Instantaneous mode.</li> <li><strong>Battery Life:</strong> 150 hours typical with continuous operation.</li> <li><strong>Power Requirement:</strong> One 9V Alkaline battery.</li> <li><strong>Operating Conditions:</strong> 0 to 55 °C, 0 to 95% RH (non-condensing).</li> <li><strong>Storage Conditions:</strong> -55 to 60 °C, 0 to 95% RH (non-condensing).</li> <li><strong>Size:</strong> 14 cm × 7.7 cm × 3.8 cm (5.5" × 3" x1.5").</li> <li><strong>Weight:</strong> 0.26 kg (0.57 lbs).</li> </ul> <p> </p> <table> <thead> <tr> <th>Sensor</th> <th>Range</th> <th>Resolution</th> </tr> </thead> <tbody> <tr> <th scope="row">Quantum</th> <td>0-199 µmol s-1 m-2<br /> 0-1999<br /> 0-19999</td> <td>0.01 µmol s-1 m-2<br /> 0.1<br /> 1</td> </tr> <tr> <th scope="row">Pyranometer</th> <td>0-19 W m-2<br /> 0-199<br /> 0-1999</td> <td>0.001 W m-2<br /> 0.01<br /> 0.1</td> </tr> <tr> <th scope="row">Photometric</th> <td>0-1999 lux<br /> 0-19999 lux<br /> 0-199 klux</td> <td>0.1 lux<br /> 1<br /> 0.01 klux</td> </tr> </tbody> </table> |
| Number of Samples per Cycle | 0 |
| Number of Samples per Day | 0 |
| Total Usage Hours per Day | 0 |
Login required. Sign in to rate this instrument and share your feedback.
Login to Give FeedbackCare and maintenance guidance for this instrument will appear here.
Daily handling, cleaning and storage guidance.
Periodic checks, calibration and servicing guidance.
Precautions and safe operating practices.
Care instructions are not available yet. Please contact the institute for guidance.