Technical Specification |
Earth voltage measurement |
|
Measuring range |
0 - 200 V AC (50,60Hz) |
Display range |
0.0 - 199.9 V |
Accuracy |
±1%rdg±4dgt |
Overload protection |
Earth resistance : 280V AC (10 seconds)
Earth voltage : 300V AC (1 minute) |
Applicable Standards |
IEC 61010-1 CAT Ⅲ300 V Pollution degree 2
IEC 61010-2-030, IEC 61557-1, -5, IEC 60529 (IP54) |
Power source |
R6(AA)(1.5V) × 6 |
Dimensions |
105(L) × 158(W) × 70(D) mm (including case lid) |
Weight |
Approx. 550g (including batteries and case lid) |
Accessories |
7095A (Earth resistance test leads [red-20m, yellow-10m, green-5m/1set])
7127B (Simplified measurement probe),
8032 (Auxiliary earth spikes[2 spikes/1set]),
9121 (shoulder strap),
9084 (Soft case)
R6(AA) × 6, Instruction manual |
Optional |
7245A (Precision measurement cord set)
8259 (Adapter for measurement terminal)
9165 (Hard case) |
|