Ultraviolet-Visible(UV-Vis) spectroscopy is a technique used to quantify the light that is absorbed and scattered by a sample (a quantity known as the extinction, which is defined as the sum of absorbed and scattered light). In its simplest form, a sample is placed between a light source and a photodetector, and the intensity of a beam of light is measured before and after passing through the sample. These measurements are compared at each wavelength to quantify the sample’s wavelength-dependent extinction spectrum. The data is typically plotted absorbance as a function of wavelength. Each spectrum is background corrected using a “blank” (a cuvette filled with only the dispersing medium) to guarantee that spectral features from the solvent are not included in the sample spectrum.
Our UV-Vis analysis is performed with Thermo Scientific GENESYS 10S Series UV-Vis spectrophotometer loaded with a high-intensity xenon lamp and dual-beam optical geometry. It can collect spectra from 190-1100 nm. The xenon lamp provides strong illumination across the ultraviolet, visible, and near-infrared electromagnetic spectrum. It is recommended to have at least 1 mL of solution without any particles to collect the spectrum. The solution will be loaded into either a quartz or plastic cell with a path length of 1 cm. The measurements of the samples as well as the solvent that is used to prepare the sample will be measured with the same instrumental configuration.
The instrument has the following capabilities:
Scan speeds up to 3,600nm/minute
Have dual-beam optics for superior photometric accuracy
Capable to acquire data from the UV to the near-IR
Increase sample throughput with the integrated 6-cell changer
Thermostating options with both circulating water and Peltier cooling (At the moment we don’t have the necessary accessories)
Measure unusual or challenging samples with a variety of optional holders for test tubes, long path cuvettes, and filters